8.3 Overview of load cells and preloads
8.3.1 Strain gauge load cells
Model Weighin pan Weigh- Replacement Preload Preload Total of load Mechanical Preload reserve Cell
(mm) ing cell no. load plate weighing plate support preload switch-on zero capacity (kg)
range support pan and weighing reserve (18% of
(kg) pan weighing range)
FTB/FTC 200 x 240 mm 3 kg 21203873 0.280 kg 0.540 kg 0.820 kg 1.180 kg 0.540 kg 5 kg
21203877*
FTB/FTC 200 x 240 mm 6 kg 21203874 0.280 kg 0.540 kg 0.820 kg 3.180 kg 1.080 kg 10 kg
21203878*
FTB/FTC 240 x 350 mm 15 kg 21203885 0.840 kg 0.880 kg 1.720 kg 3.280 kg 2.700 kg 20 kg
21203889*
FTB/FTC 240 x 350 mm 35 kg 21203886 0.840 kg 0.880 kg 1.720 kg 13.280 kg 6.300 kg 50 kg
21203890*
FTB/FTC 240 x 350 mm 60 kg 21203887 0.840 kg 0.880 kg 1.720 kg 38.280 kg 10.800 kg 100 kg
21203891*
* Premium strain gauge load cells