Guidance and manufacturerās declaration - electromagnetic immunity
308nm Excimer System is intended for use in the electromagnetic environment
specified below. The customer or the user of 308nm Excimer System should assure that
it is used in such an environment.
Electromagnetic environment -
guidance
Conducted
RF
EN
61000-4-6
Radiated
RF
EN
61000-4-3
3 Vrms
150 kHz to 80
MHz
6V m in ISM
and amateur
radio bands
between
10V/m
80 MHz to 2.7
GHz
3 V
6 V m in ISM
and amateur
radio bands
between
10 V/m
Portable and mobile RF
communications equipment should be
used no closer to any part of 308nm
Excimer System, including cables, than
the recommended separation distance
calculated from the equation applies to
the frequency of the transmitter.
Recommended separation distance
800 MHz to 2.7GHz
where p is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
metres (m).b
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, a should be