91
61000-4-5
±2 kV common
mode
± 2 kV common
mode
commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power
supply input
lines
EN
61000-4-11
< 5 % UT
(>95 % dip in UT )
for 0.5 cycle
40 % UT
(60 % dip in UT )
for 5 cycles
70 % UT
(30 % dip in UT )
for 25 cycles
< 5 % UT
(>95 % dip in UT )
for 5 sec
< 5 % UT
(>95 % dip in UT )
for 0.5 cycle
40 % UT
(60 % dip in UT )
for 5 cycles
70 % UT
(30 % dip in UT )
for 25 cycles
< 5 % UT
(>95 % dip in UT )
for 5 sec
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
308nm Excimer System
requires continued
operation during power
mains interruptions, it is
recommended that 308nm
Excimer System be powered
from an uninterruptible
power supply or a battery.
Power
frequency
(50/60 Hz)
magnetic
field
EN
61000-4-8
30A/m
30A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE:UT is the a. c. mains voltage prior to application of the test level.
Guidance and manufacturers declaration - electromagnetic
immunity for equipment and system that are not