61
If the handle is skewed and not properly inserted (as shown below),
calibration will fail or be inaccurate after the calibration is initiated.
Figure 4-16
If the prompt “Calibration Failed” is displayed, adjust the handle
position as shown in Figure 4-14 to restart the calibration procedure.
The result of the calibration performed without the handle being
properly inserted, is not accurate, even if the calibration is passed.
Readjust the handle position according to Figure 4-14 and perform the
calibration procedure again.
If the treatment handle has been inserted correctly and the