74
8
Shows "The lamp
ballast system is
overheated
abnormally"
The temperature
of the lamp ballast
is too high.
Shut down and cool down for
a while before reusing.
9
Shows "The
temperature sensor
value is incorrect"
Software failures
Shut down and restart
The cables of the
handle are not
connected
Reconnect the cables of the
handle after shutting down
the equipment
Temperature
sensor failure
Please contact the
manufacturer.
10
Shows "Calibration
failed! The light source
is low, please change
the light source
immediately".
The handle is not
placed correctly
Reload the handle
Light source
intensity is too
low
Please contact the
manufacturer for light source
replacement.
11
Stop at the boot
screen after power on
Micro SD card is
not installed.
Reinstall the Micro SD card
after shutting down.