EasyManua.ls Logo

Keysight Technologies 34970A - Page 111

Keysight Technologies 34970A
197 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Procedures 4
Keysight 34970A/34972A Service Guide 111
17 Ch 31 & 41 P3 P3 P4 P4 ________Ohms 2.00 W K301
18 Ch 32 & 42 P3 P3 P4 P4 ________Ohms 2.00 W K302
19 Ch 33 & 43 P3 P3 P4 P4 ________Ohms 2.00 W K303
20 Ch 34 & 44 P3 P3 P4 P4 ________Ohms 2.00 W K304
21 Ch 35 & 45 P3 P3 P4 P4 ________Ohms 2.00 W K305
22 Ch 36 & 46 P3 P3 P4 P4 ________Ohms 2.00 W K306
23 Ch 37 & 47 P3 P3 P4 P4 ________Ohms 2.00 W K307
24 Ch 38 & 48 P3 P3 P4 P4 ________Ohms 2.00 W K308
25 Ch 41 & 11 P4 P4 P1 P1 ________Ohms 2.00 W K401
26 Ch 42 & 12 P4 P4 P1 P1 ________Ohms 2.00 W K402
27 Ch 43 & 13 P4 P4 P1 P1 ________Ohms 2.00 W K403
28 Ch 44 & 14 P4 P4 P1 P1 ________Ohms 2.00 W K404
29 Ch 45 & 15 P4 P4 P1 P1 ________Ohms 2.00 W K405
30 Ch 46 & 16 P4 P4 P1 P1 ________Ohms 2.00 W K406
31 Ch 47 & 17 P4 P4 P1 P1 ________Ohms 2.00 W K407
32 Ch 48 & 18 P4 P4 P1 P1 ________Ohms 2.00 W K408
[a] Only the channel currently under test should be closed at one time. All other channels should be open.
Test #
Channel
Closed
[a]
External DMM Ohmmeter Connections
Measured Value
Test
Limit
Relay Measured
HI LO HI Sense LO Sense

Table of Contents

Related product manuals