Product Declassification and Security Product Name: Network Analyzer
Model Number(s): E5061B/E5063A
Page 14 of 16
Storage disk re-installation procedure:
Once the E5061B/E5063A is returned to the secured area, follow the steps listed below.
Any servicing of the E5061B/E5063A may include the regeneration of correction
constants.
1. Remove the spare “Unsecured” hard drive and replace with the original “Secured hard
drive.
2. Connect the external keyboard and mouse to the connectors on the E5061B/E5063A.
Turn on the E5061B/E5063A.
3. Press [Macro Setup] and press Load Project... in the softkey menu.
4. A dialog box will appear for you to select the program to load. Select
RestoreSysCorFile.vba from the D:\Agilent\Service folder and then press the Open button.
5. Press [Macro Run]. The RestoreSysCorFile dialog box will appear. Click OK.
6. Copy the gen.lic files from your USB memory into the E:\License directory. Manually
create the License directory if the directory does not exist.
Note: If your secured hard disk does not contain the “RestoreSysCorFile.vba” program,
copy the program from the unsecured hard disk.