EasyManua.ls Logo

Keysight Technologies 5061B - Memory Clearing, Sanitization And;Or Removal

Keysight Technologies 5061B
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Product Declassification and Security Product Name: Network Analyzer
Model Number(s): E5061B/E5063A
Page 8 of 16
Memory Clearing, Sanitization and/or Removal Procedures
This section explains how to clear, sanitize, and remove memory from your instrument for all memory types.
<Memory type>
Description and
purpose
Main Memory for Windows Operating system memory
Size
Windows XP: 2 GB, Windows 7: 4 GB, Windows 10: 4 GB
Memory clearing
Power rebooting. This is a volatile memory.
Memory sanitization
Power rebooting. This is a volatile memory.
Memory removal
This memory cannot be removed without damaging the instrument
Write protecting
N/A
Memory validation
N/A
Remarks
Description and
purpose
Media Storage (Hard Disk Drive/Solid State Drive)
Size
Windows XP/HDD:160 GB/ 320 GB, Windows 7/SSD: 80 GB/ 150 GB,
Windows 10/SSD: 150GB.
E5061B.
(Hard Disk Drive)
160 GB [Revision Prefix: ELxxx, KY203 - KY221]
320 GB [Revision Prefix: KY222]
(Solid State Drive)
80 GB [Revision Prefix: KWxxx, YD472 - YD486]
150 GB [Revision Prefix: YD488, YF]
E5063A
(Solid State Drive)
80 GB [Revision Prefix: KA, KD, KE]
150 GB [Revision Prefix: KE, FK]
Memory clearing
N/A
Memory sanitization
N/A
Memory removal
The storage drive needs to be removed and replaced with a new or unused
hard disk drive part as per the service manual.
Write protecting
N/A
Memory validation
N/A
Remarks
Description and
purpose
Memory for DSP (RAM) for A51 DSP Module
Size
1.8M bit
Memory clearing
Power rebooting. This is a volatile memory.
Memory sanitization
Power rebooting. This is a volatile memory.
Memory removal
This memory cannot be removed without damaging the instrument.
Write protecting
N/A
Memory validation
N/A
Remarks