Product Declassification and Security Product Name: Network Analyzer
Model Number(s): E5061B/E5063A
Page 5 of 16
Instrument Memory
This section contains information on the types of memory available in your instrument. It explains the size of
memory, how it is used, its location, volatility, and the sanitization procedure.
Summary of instrument memory - base instrument
Memory Type and Size
Writable During
Normal Operation?
Data Retained When
Powered Off?
Purpose/Cont
ents
Data Input
Method
Location in
Instrument and
Remarks
Sanitization
Procedure
Main Memory (RAM)
E5061B
2GB[SN:MY491xxxxx]
2GB[SN:MY492xxxxx]
4GB[SN:MY493xxxxx]
4GB[SN:MY494xxxxx]
4GB[SN:MY495xxxxx]
4GB[SN:MY496xxxxx]
4GB[SN:MY497xxxxx]
4GB[SN:MY498xxxxx]
E5063A
4GB[SN:MY541xxxxx]
4GB[SN:MY542xxxxx]
4GB[SN:MY543xxxxx]
4GB[SN:MY544xxxxx]
4GB[SN:MY545xxxxx]
Yes
No
Windows
Operating
system
memory
Operating
system (not
user defined)
E5061B
A50/A60/A70/A80/
A53 CPU Module
E5063A
A80/ A53 CPU
Module
Cycle power
Media Storage
E5061B
(Hard Disk Drive)
160 GB
[Revision Prefix:
ELxxx, KY203 -
KY221]
320 GB
[Revision Prefix:
KY222]
(Solid State Drive)
80 GB
[Revision Prefix:
KWxxx, YD472 -
YD486]
Yes
Yes
Windows
Operating
system boot
device,
factory
correction
data, and
users file
including
saved traces
data,
settings, or
images.
User-Saved
Data
Operating
system (not
user defined)
HDD/SSD assembly