2Service
148 Keysight 53210A/53220A/53230A Assembly Level Service Guide
The current internal ambient temperature and the temperature at the last
Auto-Cal can be seen simultaneously using:
[Utility] / (Cal AutoCal) / (Measure Temp)
To run auto calibration:
1 Press the Utility key and then the Cal / AutoCal softkey to display the calibrate
options.
2 Press Auto Cal.
Display stops updating until auto-calibration is completed (several seconds).
If the calibration is successful, the screen displays “Auto Calibration succeeded”.
Cal Count is not updated, but the calibration constants ARE updated.
If any one of the calibration routines fails, the screen displays “Auto Calibration
failed”.
NOTE
It is recommended that Auto-Cal be performed frequently during routine counter
operation, especially if there has been an internal ambient temperature change
of 5
o
C or more since the last Auto-Cal. A good practice is to perform Auto-Cal at
the beginning of a new day of use, at an operator shift change, or other strategic
time. However, it may not be desirable to perform an Auto-Cal in the middle of a
series of measurements, since the cal constants are updated each time Auto-Cal
is performed.