2Service
176 Keysight 53210A/53220A/53230A Assembly Level Service Guide
Device-specific errors (-300..., +800)
These error messages can be either operator information only, or can indicate a
possible hardware failure.
-310, “System error; internal software error”
(1) P0500-63001 Processor board
-310, “System error; software initialization failed”
(1) P0500-63001 Processor board
-310, “System error; out of memory”
(1) P0500-63001 Processor board
-311, “Internal software error”
-313, “Calibration memory lost”
(1) P0500-63001 Processor board
-313, “Calibration memory lost; memory corruption detected
(1) P0500-63001 Processor board
-313, “Calibration memory lost; due to firmware revision change”
-314, “Save/recall memory lost; memory corruption detected”
(1) P0500-63001 Processor board
-314,Save/recall memory lost; due to firmware revision change
-315, “Configuration memory lost; memory corruption detected”
(1) P0500-63001 Processor board
-315, “Configuration memory lost; due to firmware revision change”
-330, “Self-test failed”
-350, “Error queue overflow”
+800, “Nonvolatile memory write failure”
P0500-63001 processor board