4 Retrofitting Options
238 Keysight 53210A/53220A/53230A Assembly Level Service Guide
Figure 4-19 Connector-adapter parts
12 Pull up slightly on the "LIFT" tab on the washer-plate so that the tab blocks the
hex nut from coming unscrewed after installation.
13 Connect the supplied semi-rigid cable to the SMA connector on the
connector-adapter. Finger-tighten the connector so that the cable is
positioned to connect to the Channel 3 assembly as shown in Figure 4-16 and
Figure 4-20. Do not tighten the nut with a wrench yet.
LIFT
TAB
CONNECTOR
ADAPTER
14mm
HEX NUT
WASHER-
PLATE