EasyManua.ls Logo

Keysight Technologies 53220A - Table 1-2 the Keysight 53210 A;53220 A;53230 A Performance; Test 1: (Q) Absolute Time Base

Keysight Technologies 53220A
273 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 Performance Tests
38 Keysight 53210A/53220A/53230A Assembly Level Service Guide
Test 1: (Q) Absolute time base
This test uses the Keysight 53210A/53220A/53230A Frequency Counters to
measure the absolute frequency of the internal 10 MHz clock.
Equipment
FS725 Rubidium timebase
CG635 signal generator (E8257D-520 is a valid substitute)
Keysight 10503 BNC cables (2)
CG635 signal generator setup
1 Set Output Channel: Q
2 Set Output Level: +7 dBm
3 Set Frequency: 10 MHz
Table 1-2 The Keysight 53210A/53220A/53230A performance tests
[a]
Test description
[a] Other counter measurement functions are mathematically derived by the processor from the
parameters verified by these performance tests.
Test 1: Absolute time base
Test 2: Frequency accuracy - channels 1 and 2
Test 3: Single period
Test 4: Pulse width
Test 5: Time interval
Test 6: Frequency accuracy - microwave channel options 106 and 115
Test 7: Pulse burst frequency - 53230A channel 3 option 150

Table of Contents

Related product manuals