EasyManua.ls Logo

Keysight Technologies 53220A - Page 216

Keysight Technologies 53220A
273 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Retrofitting Options
216 Keysight 53210A/53220A/53230A Assembly Level Service Guide
5 Remove the rear panel plug(s) for the Channel(s) being installed.
6 Insert the BNC connector end of the RF cable (over the top of the Option 106/
115 assembly) through the appropriate channel cutout hole in the rear panel.
7 Using a 14 mm deep-socket or spintite, secure the connector with the
included hex nut and lockwasher (do not overtighten).
8 If necessary, repeat steps 2 through 7 for Channel 2.
9 Install the supplied cable clamp(s) as shown in Figure 4-7.
10 If the Option 300 Internal Battery was removed at the beginning of this
procedure, re-install it now. Be sure to re-connect the cables.
11 Re-install the cover by performing the cover removal procedure (Preliminary
Steps 1-4), in reverse.
NOTE
When Option 201 is installed (Channels 1 and 2 rear panel BNC connectors), the
front panel connector performance for RF Channels 1 and 2 is not specified. The
rear panel connector for each channel IS specified according to the information
in http://literature.cdn.keysight.com/litweb/pdf/5990-6283EN.pdf.

Table of Contents

Related product manuals