Backdating 6
Keysight 53210A/53220A/53230A Assembly Level Service Guide 267
10 Press the Perform Cal softkey:
Display softkey annunciators show:
11 Press the Cal Step softkey.
Display softkey annunciators show:
* Only displayed if Option 106 or 115 is installed.
12 Press the uWave softkey.
Display softkey annunciators show:
13 Use the front panel knob and left/right arrow keys to enter the value as: 500
MHz in the data entry area to agree with your source.
14 Press the Input Power softkey and enter the power level of the 500 MHz source
as -27 dBm. (Must be -27 dBm ± 0.5 dB at the input connector. If necessary,
verify with a power meter.)
**
NOTE
Alternatively, you may use the Shift key on the front panel to put the keyboard
keys into a numeric mode, and then enter the entire frequency manually. The
data entry area will repeat whatever numbers are used to select the calibration
frequency.