1 Performance Tests
52 Keysight 53210A/53220A/53230A Assembly Level Service Guide
9 Press and set:
–Source: Manual
Wait For Trig is displayed.
–Trig Delay: 0 Seconds
–Rdgs/Trigger: 100
10 Press the Math key.
The following softkeys are displayed:
11 Press and set:
–All Math: On
12 Press the Statistics softkey:
–The Statistics softkeys are displayed: