Performance Tests 1
Keysight 53210A/53220A/53230A Assembly Level Service Guide 63
12 Press the Trigger key.
Front panel display shows:
13 Press to set:
–Source: Manual
Wait For Trig is displayed
Rdgs/Trigger: 100
–Trig Delay: 0 seconds
14 Press the Math key.
The following softkeys are displayed:
15 Press and set:
–All Math: On