Performance Tests 1
Keysight 53210A/53220A/53230A Assembly Level Service Guide 75
Channel 3 (option 115) frequency accuracy test procedure
1 Press the Trigger key.
The counter takes 100 readings and fills in the math functions.
2 Verify the ‘Mean’ is:
300 MHz ± 3.0 Hz (53210A)
300 MHz ± 0.3 Hz (53220A/53230A)
3 Verify the ‘Std Dev’ is less than:
4.2 Hz (53210A)
2.1 Hz (53220A)
0.42 Hz (53230A)
4 Record the Mean and Standard Deviation for 300 MHz, Test 6b, in the
Keysight 53210A/53220A/53230A Performance Test Record.
5 Press the Reset Stats softkey.
6 Set the E8257D-520 to 3.1 GHz at -27 dBm.
7 Press the Trigger key.
The counter takes 100 readings and fills in the math functions.
NOTE
The input power MUST be -27 ±0.5 dBm at the channel 3 input connector. If
necessary, use a 3.5 mm cable with certified loss at 3.1 GHz or a power meter to
verify input power.