16 Keysight E5080B ENA Series RF Network Analyzers
Installation
Power Supply
1-
Power Supply
Before turning on the E5080B power, check the following.
Verification of the Power Supply
Confirm that the power supplied to the E5080B meets the following
requirements:
Verification and Connection of Power Cable
The three-wire power cable attached to the E5080B has one wire serving as a
ground. Using this power cable allows the E5080B to be grounded, thereby
protecting you against electrical shock from the power outlet.
Step 1. Confirm that the power cable is not damaged.
Step 2. Use the supplied cable to connect between the power cable receptacle
(Figure 1-2 on page 19) on the rear panel of the E5080B and a three-wire
power outlet with the grounding prong firmly connected in the ground slot.
For power cord list and options, refer to www.keysight.com/find/powercords.
Requirements
Voltage 100 to 240 VAC
Frequency 50/60 Hz
Maximum power consumption 350 W
NEVER use a power cable showing any sign of damage. Faulty cables can
cause electrical shock.
Use the supplied power cable with grounding wire to securely ground the
E5080B.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5080B and is the answer not in the manual?

Keysight Technologies E5080B Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5080B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals