Keysight FieldFox Handheld Analyzers Service Guide 33
Tests and Adjustments
Before Returning an Instrument to a Customer
Before Returning an Instrument to a Customer
When finished servicing an FieldFox analyzer, perform the following before
returning the unit to the customer:
— Set the Display Colors to Indoor and the display Brightness to 80% via
System (7) > Preferences > Preferences, etc.
— Perform an Operational Check as described in the following section.
— Connect and disconnect the external battery charger while the FieldFox is
operating and confirm that the power cord icon is displayed appropriately
at the top of the display.
— Charge the battery to 80% or higher.
— Connect FieldFox to the internet.
—Press System > System Configuration > More > update time using
Internet.