174                                                                      Keysight FieldFox Handheld Analyzers Service Guide
General Purpose Maintenance and Service Procedures
Firmware Upgrades
8-
Firmware Upgrades
To Check the Current Firmware Revision
—Press System(7).
— Press Service Diagnostics.
—Press System Information.
The current firmware version is indicated on the “Firmware Release” line.
For information on upgrading FieldFox firmware, go to 
www.keysight.com/find/fieldfoxsupport.
If the firmware upgrade process is interrupted, memory contents in the 
instrument may become corrupted. If a problem is suspected, consider doing a 
fresh installation of the firmware. Follow the process under “Install Firmware 
from SD Card” near the end of Chapter 7.
See the “Firmware Related Information” section of Chapter 5 for details on 
what happens during the firmware update process.
Miscellaneous Service and Maintenance
Cleaning the Display
Materials: Use a soft, lint-free cloth. Use a solution with 99.5% IPA (isopropyl 
alcohol) or a high quality window cleaning solution.
Technique: Do not rub back and forth with a polishing motion. Use a single 
swipe from the soiled location to the nearest edge of the display.
Battery Maintenance
See the “Working with the Lithium-Ion Battery” section of the FieldFox User’s 
Guide for information on the removable battery. In particular, see “Built-In 
Battery Gauge” and “Reconditioning the Battery”.