24 Keysight M9036A Embedded Controller Startup Guide
Step 6: Install User Application Modules in the PXI Express Chassis
Step 6: Install User Application Modules in the PXI Express Chassis
The PXIe chassis accepts modules conforming to the PXI standard, which can be
installed in available slots in the chassis. Turn off the chassis prior to installing
modules. These may include:
- PXI instrument or switch modules
- PXIe filler panel modules (PXIe chassis are shipped with filler panel modules in
all but Slot 1; you may need to remove filler panels from covered slots.)
Install application module drivers (if necessary)
Installed application modules will typically require device drivers and control
software. You need not install them to verify basic chassis or embedded
controller operation, but would logically install them on the controller at this
time. For others, consult the module provider for software requirements and
instructions.
- Static Electricity—The components and connectors on
modules are sensitive to static electricity. To minimize
electrostatic damage, take the necessary anti-static
precautions. Both chassis provide a grounding terminal, to
which you can connect a wrist strap.
- Empty Slots—Except for performing initial chassis verification
or troubleshooting, do not operate the chassis with empty
slots. Always insert a filler panel or instrument module into
empty slots. This is especially important for the slots on either
side of an instrument module. This allows proper air flow and
cooling, and provides EMI shielding for the chassis and
installed components. Leaving slots empty can increase fan
speed, raise ambient noise, overheat components, and shut
down modules.
- Hot Swap—PXIe does not explicitly support hot swap for
instrument modules. Keysight recommends fully powering
down the chassis before installing or removing modules
You may need to connect the embedded controller to the internet
to locate and install the necessary drivers.

Other manuals for Keysight Technologies M9036A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies M9036A and is the answer not in the manual?

Keysight Technologies M9036A Specifications

General IconGeneral
BrandKeysight Technologies
ModelM9036A
CategoryController
LanguageEnglish

Related product manuals