EasyManua.ls Logo

Keysight Technologies M9347AH02 - Frequency, Power, and Phase Settings

Keysight Technologies M9347AH02
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 M9347A SCPI Command Reference
IVI Equivalent
KtM9347.Synthesizers[“1”].PowerLevel = -5.0;
Phase Offset ([:DDS[1]|2])
[:DDS[1]|2]:PHASe <val>
[:DDS[1]|2]:PHASe? [MAXimum|MINimum]
This command applies a phase rotation, in degrees, to the data generated by the
synthesizer.
The query returns the maximum or minimum allowable phase rotation if the optional
MAXimum or MINimum are used.
Example
:DDS2:PHAS -45
The preceding example applies a -45° phase rotation to synthesizer channel #2.
IVI Equivalent
KtM9347.Synthesizers[“2”].PhaseOffset = -45.0;
Output Enabled ([:DDS[1]|2])
[:DDS[1]|2]:OUTPut[:STATe] ON|OFF|1|0
[:DDS[1]|2]: OUTPut[:STATe]?
This command enables or disables the RF output of the synthesizer.
Example
:DDS2:OUTP OFF
The preceding example disables the RF output of synthesizer channel #2.
IVI Equivalent
KtM9347.Synthesizers[“2”].OutputEnabled = false;
Calibration Subsystem (:CALibration)
System Clock Calibration
:CALibration:SCLK
This command invokes an internal alignment to lock the DDS chips to the system
clock. This alignment must be performed whenever the system clock source is
changed.
Example
:CAL:SCLK
IVI Equivalent
KtM9347.Calibration.Align( AlignmentType.SystemClock );
12 Keysight M9347AH02 Getting Started Guide