EasyManua.ls Logo

Keysight Technologies M9370A

Keysight Technologies M9370A
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-44 Keysight M937xA PXIe Service Guide
Tests, Adjustments, and Troubleshooting
Performance Tests
2-
7. The test set’s output attenuator is changed in 5 dB steps from 0 to 60 dB,
and this measurement is repeated.
8. With the 1 dB of overlap in each measurement, the data for each attenuator
setting can be stitched together to provide a complete receiver linearity
profile from +3 dBm to –63 dBm.
9. This test is repeated for each receiver.
If the Analyzer Fails this Test
- If the analyzer still fails this test, repeat the test.
- If the analyzer fails this test, return it to Keysight for repair. Refer to
“Shipping Your Network Analyzer for Service or Repair” on page 1-3.

Table of Contents

Other manuals for Keysight Technologies M9370A

Related product manuals