2-46 Keysight M937xA PXIe Service Guide
Tests, Adjustments, and Troubleshooting
Adjustments
2-
Source Adjustment
The source calibration is used to adjust your network analyzer for a flat source
power across its full frequency range. There are differences between each test
port; therefore, an adjustment is required for each port.
Equipment Used for the Source Adjustment
Procedure
1. Connect the USB power sensor and adapter to the analyzer Port 1. If using
a power meter, make the proper GPIB connections. Refer to “Get the
Hardware Required for your GPIB Instruments on page 2-5.
2. Press Utility > System > Service > Adjustments.
3. In the Select Desired Adjustment dialog box, click Source Adjustment.
4. Follow the instructions and prompts as they are displayed. If using a power
meter, ensure the GPIB settings for the sensor(s) are correct in the dialog
box.
Equipment Type Model or Part Number Alternate Model or Part Number
USB Thermocouple power sensor U8485A Option 200
N8485A power meter
a
Adapter, 3.5 mm -m- to 3.5 mm
-m-
83059A 85052-60014
a. Refer to “Get the Hardware Required for your GPIB Instruments” on page 2-5.

Table of Contents

Other manuals for Keysight Technologies M9370A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies M9370A and is the answer not in the manual?

Keysight Technologies M9370A Specifications

General IconGeneral
BrandKeysight Technologies
ModelM9370A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals