EasyManua.ls Logo

Keysight Technologies N9000B - Benchtop Configuration Disassembly

Keysight Technologies N9000B
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Installation Note N9000-90039
Option T03 and T06, Tracking Generator Upgrade Kit
Standard Instrument (Benchtop Configuration)
1. Disconnect the instrument from ac power.
2. Refer to Figure 1. Using the T-20 driver, remove the four screws (two on each side) (1) that
attach the handle strap (2) on each side of the instrument.
3. Using the T-20 driver, remove the four screws (including washers) (3) that hold the rear feet
(4) in place. The rear feet will not be re-used.
4. Pull the instrument cover (5) off towards the rear of the instrument. The instrument cover will
not be re-used.
Figure 1 Standard Instrument Outer Case Removal
5. Proceed to the Front Frame Assembly Removal section to remove the front frame.

Other manuals for Keysight Technologies N9000B

Related product manuals