10 Installation Note N9040-90014
Option RTS, Streaming
Alternate Installation Procedure
The License File can be manually installed over USB or LAN by placing the license file in the
following folder on the signal analyzer
C:\Program Files\Agilent\licensing
Verify the License Installation and Hardware
1. Cycle power on the signal analyzer and wait until the analyzer boots to the measurement
application screen.
2. Press System, Show System to display a list of installed options.
3. Verify that the installed options list contains the newly installed N9040B-RTS.
4. Press System, Show Hard ware. There will be an entry for the WBDIF Extension.
5. Press MODE/MEAS, choose Real-Time Spectrum Analyzer. Press OK.
6. Press Input/Output, Output, Wideband Digital BUS ON. Look at the rear panel streaming
connectors, and assure the green Streaming LED is on.
Utilities, Adjustments, and Performance Verification Tests
Utilities Required
None
Adjustments Required
Performance Testing Required
For assistance, contact your nearest Keysight Technologies Sales and Service Office.
To find your local Keysight office access the following URL, or if in the United States, call the
following telephone number:
http://www.keysight.com/find/assist
1-800-829-4444 (8 am - 8 pm ET, Monday - Friday)