[SENSe:]CAPacitance:NULL:VALue {<value>|MIN|MAX|DEF}
[SENSe:]CAPacitance:NULL:VALue? [{MIN|MAX|DEF}]
Stores a null value for capacitance measurements. The null value is subtracted from each sample if the
null state is turned on.
Parameter Typical Return
<value> -120 to +120 µF. Default: 0. -1.20000000E-04
Configure capacitance measurements, turn null function on. Make two measurements and send them to
the instrument's output buffer:
CONF:CAP AUTO
CAP:NULL ON
CAP:NULL:VAL 1.2E-12
SAMP:COUN 2
READ?
Typical Response: +3.01963841E-10,+3.01566277E-10
l Specifying a null value disables automatic null value selection
([SENSe:]CAPacitance:NULL:VALue:AUTOOFF).
l To use the null value, turn on the null state with the [Sense:]CAPacitance:NULL[:STATe] command.
l This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset
(SYSTem:PRESet), or CONFigure function.
Keysight Truevolt Series Operating and Service Guide
315
SCPI Programming Reference