5 Triggering and Gating
172 Keysight 53220A/53230A User’s Guide
Setting the gate time
A measurement gate time can be specified directly with the command: 
[SENSe:]FREQuency:GATE:TIME {<time>|MINimum|MAXimum| 
DEFault}
[SENSe:]FREQuency:GATE:TIME? {MINimum|MAXimum|DEFault}
(query form)
CONFigure and MEASure automatically set the gate source to TIME, and the gate 
time according to their optional expected value and resolution parameters. 
Specifying these parameters or using default values with these commands 
simplifies counter programming. However, setting/changing the gate time 
separate from the CONFigure/MEASure command allows you to change only the 
gate time while preserving other configuration settings.
Querying the gate time The gate time set by the expected value and resolution 
parameters can be determined after execution of the CONFigure or 
MEASure command. 
For example, assume a measurement requiring 10-digit (mHz) resolution of a 
5 MHz signal is required. Because digits of resolution is approximately equal to 
Log
10
 (expected value) - Log
10
 (resolution), this measurement could be 
configured as:
MEAS:FREQ? 5e6, 5E-4, (@1) // expected freq, resolution
A typical measurement using this configuration might return:
+4.99998458333282E+006
With the counter display registering:
4.999 984 583 3MHz (11 digits - resolution enhanced)