6 53220A/53230A Math, Graphing, and Data Logging
246 Keysight 53220A/53230A User’s Guide
Graphics Functions and Reading Memory
Figure 6-13 summarizes the effects on reading memory when statistics and
histograms are reset, and when trend charts are cleared.
Figure 6-13 Reading memory and resetting/clearing graphics.
reading memory
.
.
.
.
.
.
(record)
.
.
.
(record)
(record)
Rese t Hi st ogr a m:
- clears current histogram (record)
- does not clear reading memory, does not abort readings
– clears count and new histogram re-starts on next reading
after reset
(record)
Clear Readings ( Trend Chart) :
- clears the trend chart
- clears reading memory, does not abort readings
- clears count and re-starts on next reading after clear
Reset St at s:
- clears current set of statistics (record)
- does not clear reading memory, does not abort readings
– clears count and new set of statistics re-start on next
reading after reset
(memory)