53220A/53230A Math, Graphing, and Data Logging 6
Keysight 53220A/53230A User’s Guide 211
Reference example
//determine the percentage difference of 100 frequency
//measurements to a reference value of 50000.000
CONF:FREQ 50E3,(@1)
SAMP:COUN 100 // take 100 readings
CALC:STAT ON // enable CALCulate1 subsystem
CALC:SCAL:STAT ON // enable scaling
CALC:SCAL:FUNC PCT // select scale function (PCT)
CALC:SCAL:REF 50.000E3 // set reference value
INIT
Scale gain and offset
The function CALCulate[1]:SCALe:FUNCtion SCALe performs an Mx-B
operation on each reading where M is the gain value, x is the reading, and B is the
offset. The gain value used in the equation is set with the command:
CALCulate[1]:SCALe:GAIN {<gain>|MINimum|MAXimum|DEFault}
CALCulate[1]:SCALe:GAIN? [{MINimum|MAXimum|DEFault}
(query form)
gain values can range from:
-1.0E+15 to -1.0E-15, 0.0, +1.0E-15 to +1.0E+15
The default gain is 1.0 and is the value set following a reset (*RST) or instrument
preset (SYSTem:PRESet).
The offset value (B) is set with the command:
CALCulate[1]:SCALe:OFFSet {<offset>|MINimum|MAXimum| DEFault}
CALCulate[1]:SCALe:OFFSet? [{MINimum|MAXimum|DEFault}]
(query form)
offset values can range from
-1.0E+15 to -1.0E-15, 0.0, +1.0E-15 to +1.0E+15
The default offset is 0.0 and is the value set following a reset (*RST) or instrument
preset (SYSTem:PRESet).