Formats and Data Flow 7
Keysight 53220A/53230A User’s Guide 269
Specifying power-on states from the front panel
A specific power-on state can be selected from the front panel as follows:
1 From the Store/Recall menu, select User Def using the ‘Power On’ soft key.
2 Press User Def File to view the file system. Select the desired state to be set at
power-on by selecting the desired folder and file name. Remember that
instrument state locations 0-4 are identified by file name (e.g. STATE_3.sta).
For states saved in other files (locations), specify the appropriate file with the
.sta extension.
Selecting Last using the ‘Power On’ soft key will recall the instrument state at the
time the counter was last turned off (state location 0). With Factory selected, the
counter will power-up in its factory-set state.