A 53220A/53230A Error Messages
294 Keysight 53220A/53230A User’s Guide
+800
Non-volatile memory write
failure
The user preferences (.prf) file is invalid and could not be stored in
internal flash memory or on the USB memory device.
+810 State has not been stored
The instrument state file specified by the MMEMory:LOAD:STATe
<“file”> contains an invalid instrument state and cannot be loaded.
+820
Model and Serial Numbers not
restored
Following replacement of the P500 processor (controller) board or main
measurement board, the model number and/or serial number do not
match or have not been re-stored. Re-store the model number and
serial number as prompted from the front panel. Refer to the service
guide for additional information.
+821
Controller and measurement
board model numbers do not
match
+822
Controller and measurement
board serial numbers do not
match
Calibration errors
+701
Calibration error;
security defeated
Shorting the calibration secure jumper (CAL ENABLE) when turning
ON the instrument will cause this error to occur indicating the security
password has been overwritten. See “Resetting the Security Code to a
Null” in the service guide for details.
+702
Calibration error;
calibration memory is secured
A calibration cannot be performed when calibration memory is secured.
See “To Unlock The Counter For Calibration” and “To Lock the Counter
in Chapter 2 of the service guide for more information. Use the
CAL:SEC:STAT ON command to enter the security code using the
remote interface.
+703
Calibration error;
secure code provided was
invalid
The security code specified was invalid.
+704
Calibration error;
secure code too long
The security code specified is greater than 12 characters.
+705
Calibration error;
calibration aborted
An in-progress calibration was aborted (stopped) by the user.
+706
Calibration error;
provided value out of range
A calibration value was entered that was outside the expected range of
valid inputs.