Disturbance Analyzer (Click) Measurements
Making a Measurement
Making a Measurement
Setting up a Click measurement
Step Action Notes
1 Turn on the receiver a. Press the front-panel power key.
2 Test Set Up a. Configure the Device Under Test
(DUT)
b. Connect the DUT power cable to
a Line Impedance Stabilization
Network (LISN)
c. Connect the LISN to the mains
power
d. Connect the output RF port from
the LISN to the N9038A MXE
EMI Receiver
3 Select the
Disturbance
Analyzer
measurement
e. Press Mode, EMI Receiver,
Meas, Disturbance Analyzer
(Click)
This brings up the following Click
Measurement display.