EasyManua.ls Logo

Keysight E3633A - Page 159

Keysight E3633A
205 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error Messages 5
Keysight E3633A and E3634A User’s Guide 159
-123
Numeric overflow
A numeric parameter was found whose exponent was larger than 32,000.
-124
Too many digits
A numeric parameter was found whose mantissa contained more than 255
digits, excluding leading zeros.
-128
Numeric data not allowed
A numeric parameter was received but a character string was expected.
Example:
DISP:TEXT 123
-130
[a]
Suffix error
A suffix was incorrectly specified for a numeric parameter. You may have
misspelled the suffix or the numeric parameter does not accept a suffix.
Example:
TRIG:DEL 0.5 SECS
-131
Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may have
misspelled the suffix.
Example:
TRIG:DEL 0.5 SECS
-134
Suffix too long
A suffix for a numeric parameter contained too many characters.
-138
Suffix not allowed
A suffix was received following a numeric parameter which does not accept a
suffix.
Example:
STAT:QUES:ENAB 18 SEC (SEC is not a valid suffix).
[a] This error message is only applicable for serial MY53xx6xxx.

Table of Contents