Front-Panel Operation 3
Keysight E3633A and E3634A User’s Guide 81
To set the HPIB Address
To configure the power supply for the HPIB interface, proceed as follows:
1 Turn on the remote configuration mode.
You will see the above message on the front-panel display if the power supply
has not been changed from the factory setting. If “RS-232” appears, choose
“HPIB / 488” by turning the knob to the right.
2 Move to the HPIB address setting mode.
The address is set to “05” when the power supply is shipped from the factory.
Notice that a different HPIB address may appear if the power supply has been
changed from the factory setting.
3 Turn the knob to change the HPIB address.
The displayed address is changed when turning the knob to the right or left.
4 Save the change and turn off the I/O configuration mode.
The address is stored in non-volatile memory, and does not change when
power has been off or after a remote interface reset. The power supply displays
a message to show that the change is now in effect. If the HPIB address is not
changed, “NO CHANGE” will be displayed for one second.
HPIB / 488
ADDR 05
CHANGE SAVED
NOTE
To exit the I/O configuration mode without any further changes, press the “I/O
Config” key until the “NO CHANGE” message is displayed.