3 Front-Panel Operation
60 Keysight E3633A and E3634A User’s Guide
Adjust the output voltage level
1 Lower the output voltage level.
Lower the output voltage level below the OVP trip point after pressing
key. The OVP and Limit annunciators are blinking.
2 Move to the clear mode.
Press key twice to move to the OVP CLEAR mode. The “OVP ON”
message appears on the display. Turn the knob to the right until the above
message appears on the display.
3 Clear the overvoltage cond ition and exit this menu.
Now, when you press key again, the “DONE” message is displayed for
a second and the OVP annunciator will not blink any more. The output will
return to meter mode.
OVP CLEAR