EasyManua.ls Logo

Keysight E4412A - Page 31

Keysight E4412A
43 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation and Service Guide 1
Keysight E4412A and E4413A Operating and Service Guide 31
Procedure
1 Turn on the signal generator and power meter (with the reference sensor
connected). Connect the DUT to channel A of he power meter (E4416/7A), and
channel B of the reference sensor (E4412A).
Allow them to warm up for approximately an hour.
2 Zero and calibrate both the DUT and reference sensor.
3 Connect the power splitter (11667B/11667C) to the RF output of the signal
generator (N5182A). The equipment setup is as shown in Figure 1-6.
Figure 1-6 Power linearity performance verification equipment setup
4 Set the continuous wave signal frequency of the signal generator, DUT, and
reference sensor to 50 MHz. Set the DUT to average-only mode.
5 Start tuning the signal generator and/or optionally tune the attenuator or
switch driver until the DUT (channel A) measures the power level as close as 0
dBm. Record the values as P
DUT
at 0 dBm and P
ref
at 0 dBm.
CAUTION
Do not exceed the maximum input power (27 dBm) of the power splitter to
avoid damage to the power splitter.

Related product manuals