Keysight E8257D/67D & E8663D PSG Signal Generators Installation Guide 23
Operation Verification
Checking the Output Power
2. Connect the equipment as shown:
3. Press Preset.
4. Press RF On/Off. The RF On annunciator is now displayed.
5. Press Mod On/Off. The Mod Off annunciator is now displayed.
6. Press Frequency and enter the first frequency value listed for your instrument model in Table 3-7,
Table 3-8, or Table 3-9.
NOTE Some typical option configurations and their maximum output power limits are shown in
Table 3-7 or Table 3-8. For the complete matrix of options or for the latest changes to
specifications, refer to the E8257D, or E8267D data sheets either on the web at
http:\\www.keysight.com\find\psg or on the documentation CD that was shipped with your
instrument.
7. Press Amplitude and enter the amplitude value for your option and frequency.
8. Measure the power level. Record any power levels not within the limits indicated.
9. Repeat steps 6 through 8 and measure the power levels for each remaining frequency listed in
Table 3-7, Table 3-8, or Table 3-9.