5.1 Performance Specification
5.2 Operating Environment
Storage Temperature
−40 °C to +70 °C
Operating Temperature
0 °C to 40 °C
Humidity
95%R.H. (at 40 °C)
Altitude
Up to 2000 m
Installation
Category II
Pollution
Degree 2
5.3 General
Power consumption
60 W (nom), 65 GSa/s operation
Safety designed to
IEC 61010-1, UL 61010-1, CAN/CSA-C22.2 No. 61010-1
EMC tested to
IEC61326-1
Warm-up time
30 min
Calibration interval
2 years recommended
Warranty
3 years standard
Cooling
Requirements
When operating the M8195A choose a location that provides at least 80 mm
of clearance at rear, and at least 30 mm of clearance at each side for the
AXIe chassis.