Installation Note N9020-90185 7
Option CR3 or CRP Installation Kit
Functional Check:
Within one year of the initial calibration date of the instrument this option is fully calibrated with no
further adjustment or verification testing. To ensure that this newly installed option has been
enabled correctly the following functional check is recommended.
1. Press System, Show, Errors, Clear Error Queue. This will remove out-of-date errors.
2. Press System, Alignments, Align Now, All.
3. Watch for any errors during the instrument alignment.
4. Press System, Show, Errors to view the instrument error log.
5. Verify that there were no errors during the alignment.