6-6 Installation and Service Guide N5251-90001
Performance Tests and Checks 
System Performance Verification
PNA Series Microwave Network Analyzer System 
N5251A
System Performance V
erification
Since the N5251A has typical (non-warranted) specifications only, the 
SYSTEM 
PERFORMANCE VERIFICATION is not a PASS/FAIL test. The SYSTEM 
PERFORMANCE VERIFICATION is a procedure to establish current performance 
of the system. See “Interpreting the Verification Results” on page 6-15 for more 
detail.
System verification is performed at the N5251A system 1.0 mm port connectors over a frequency 
range of 45 MHz to 110 GHz with a Keysight 85059A 1.0 mm Calibration and Verification kit. The 
basic verification process checks a pair of ports at one time. Verification of a 2-port system typically 
takes about 20 minutes. Verification of a 4-port system typically takes about 35 minutes.
The system verification procedure is automated by the analyzer firmware. For each verification 
device, the system measures the magnitude and phase for all four S-parameters and reads data for 
the device from the 85059A kit. The procedure displays the measured data along with the kit data to 
allow a visual comparison.
During the system verification process, it is NECESSARY that the verification devices be measured with 
their female connectors connected to Port 1 and their male connectors connected to Port 2. A 1.0 
mm female-to-female cable connected to Port 2. This cable is included in the 85059A 1.0 mm 
Calibration and Verification kit.
When to Verify
After installation of the system is complete, a performance verification is necessary to assure proper 
system operation. This initial verification is included with the installation.
After the initial verification, the verification should be repeated once a year. This recommended 
interval 
assumes that Keysight cables are used with the system.
If non-Keysight cables, adapters, or other fixtures are used, the verification schedule must be 
determined by the user, as the characteristics of these devices are unknown. In establishing a 
verification schedule, the following factors should be considered:
• Frequency of use
• Amount of cable movement
• Amount of drift occurring between prior verifications 
Performance verification of a system performed at long intervals is not to 
be confused with measurement calibration. Measurement calibration 
typically is performed on a daily basis, or when the measurement setup or 
conditions have changed.