6-16 Installation and Service Guide N5251-90001
PNA Series Microwave Network Analyzer System 
N5251A
Performance Tests and Checks
System Performance Verification
Improving the Verification Results 
IMPORTANT Inspect all connections. Do not remove the test port cable from the analyzer 
test port. This will invalidate the calibration that you performed earlier. 
1. Disconnect and clean the device that failed the verification measurement.
2. Reconnect the device making sure that all connections are carefully torqued.
3. Measure the device again using the previous calibration.
4. If the device still fails the verification measurement, repeat the full system verification process 
including calibration. 
Saving Verification Results
When you have system verification results that represent current system performance, save the 
results files (five files for each port pair) in the appropriate subdirectory under D:\sysver results.