Language Reference 5
Series N8700 User’s Guide 75
Source Commands
Source commands program the voltage, current, triggered, and
protection functions.
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <value>|MIN|MAX
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <value>|MIN|MAX
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX]
These commands set the immediate and the triggered output current.
Values are programmed in amperes. The immediate level is the
output current setting. The triggered level is a stored value that
transfers to the output when a trigger occurs. The *RST values = Min.
Model ( I rating) 400A 330A 220A 165A 110A 85A 55A 42A 33A 22A 11A 5.5A
Minimum current 0 0 0 0 0 0 0 0 0 0 0 0
Maximum current 420 346.5 231 173.25 115.5 89.25 57.75 44.1 34.65 23.1 11.55 5.775
Model ( I rating) 250A 170A 125A 85A 65A 50A 34A 17A 8.5A
Minimum current 0 0 0 0 0 0 0 0 0
Maximum current 262.5 178.5 131.25 89.25 68.25 52.5 35.7 17.85 8.925
[SOURce:]CURRent:PROTection:STATe ON|OFF
[SOURce:]CURRent:PROTection:STATe?
This command enables or disables the over-current protection (OCP)
function. The enabled state is On (1); the disabled state is Off (0). If
the over-current protection function is enabled and the output goes
into constant current operation, the output is disabled and OC is set
in the Questionable Condition status register. The *RST value = Off.
An over-current condition can be cleared with the Output Protection
Clear command after the cause of the condition is removed.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <value>|MIN|MAX
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <value>|MIN|MAX
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX]
These commands set the immediate and the triggered output voltage.
Values are programmed in volts. The immediate level is the output
voltage setting. The triggered level is a stored value that transfers to
the output when a trigger occurs. The *RST values = Min.
The range of values that can be programmed for these commands is
coupled with the voltage protection and the voltage limit low settings.
The maximum value for the immediate and triggered voltage level is
either the value in the following table, or the voltage protection
setting divided by 1.05; whichever is lower. The minimum value is
either the value in the table, or the low voltage setting divided by
0.95; whichever is higher.