Installation Note N9048-90017 17
Option EXM, External Mixing Upgrade Kit
3. Refer to Figure 13. Remove the 50 Ω loads on J8 and J13 if present. Connect the end of W30
labeled “13” to A13A1J13. Torque the cable nut to 10 inch-pounds. Route the cable as shown.
Figure 13 W17 and W30
4. Refer to Figure 13. Locate semi-rigid coax cable, part number N9020-20167 in the upgrade kit.
This is W17. Connect the end with the SMA male connector A13A1J8, with the SMA female
connector pointing towards where the front panel would be. The long, straight section of W17
should be parallel to the casting of the A13 Front End and level. Torque the cable nut to
10 inch-pounds.