24 Installation Note N9048-90017
Option EXM, External Mixing Upgrade Kit
5. Refer to Figure 21. Align and connect cables W17 and W29. Hand-tighten the nut.
Figure 21 W17 and W29 Properly Aligned
6. Refer to Figure 22. Use a ¼” open-end wrench to prevent the SMA female connector on W17
from twisting. Use a 5/16” torque wrench to torque the nut on W29 to 10 inch-pounds.