Installation Note N9030-90058 3
Option EXM, External Mixing Upgrade Kit
Option EXM, External Mixing Upgrade Kit
Software and test equipment is required for making adjustments and for performance verification
testing.
Information on how to obtain this software can be found at:
www.agilent.com/find/calibrationsoftware
While Agilent does recommend that a full calibration be performed after the installation of this
upgrade, the end user must ultimately determine whether they want this service or not. If a full
calibration is required, arrangements regarding the level of calibration must be made between the end
user and the calibration provider.
NOTE Option EXM requires instrument software A.10.00 or later. The latest revision of
the analyzer software may be downloaded from:
http://www.agilent.com/find/xseries_software
Products Affected: PXA N9030A
Serial Numbers: US/MY/SG51160000 and above
(Options 543, 544, and 550)
US/MY/SG51380000 and above
(Options 503, 508, 513, and 526)
To Be Performed By: (X) Agilent Service Center
(X) Personnel Qualified by Agilent
( ) Customer
Estimated Installation Time:
Estimated Adjustment Time:
Estimated Verification Time:
1.5 Hours
1.0 Hours
a
4.0 Hours
a
a. To ensure that these newly installed options are functioning properly, the procedure that follows
includes the requirement of performing certain adjustments and performance verification tests.
However, the completion of these tests does not guarantee that the instrument meets all advertised
specifications.