10 U2722A/U2723A Service Guide
2Getting Started 
 
Introduction
Self-Test 
• A brief power-on self-test occurs automatically whenever 
you turn on the instrument. This limited test assures that 
the instrument is capable of operation.
• If the self-test fails, the power indicator on the front 
panel will start blinking. You can use the SYSTem:ERRor? 
query command from the remote interface to read the 
error messages. If repair is required, contact a Keysight 
Service Center. Please refer to the programmer’s reference 
guide for more information on the error messages.
• Self-test may be initiated remotely by sending the *TST? 
SCPI command to the instrument.
Keysight Measurement Manager
To start the self- test procedure using the Keysight 
Measurement Manager, perform the following steps.
1 Click  Tools > Self-Test.
Disconnect all connections from the output terminals before 
performing the self-test procedure. Ensure that the output terminals 
are not connected to any load before you turn on the U2722A/U2723A.