EasyManua.ls Logo

Keysight u2722a - Source Current Mode Programming and Readback Accuracy

Keysight u2722a
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Verification and Adjustment 3
U2722A/U2723A Service Guide 25
Source Current Mode Programming and Readback Accuracy
Table 3 - 5 Source current mode programming and readback accuracy
Iteration Test point Current range Clamp voltage Voltage range I
meas
– I
drive
I
rdbk
– I
meas
Limit
1
0.5 μA1 μA
20 V 20 V ±1.275 nA
2 2 V 2 V ±1.275 nA
3
5 μA 10 μA
20 V 20 V ±12.75 nA
4 2 V 2 V ±12.75 nA
5
50 μA100 μA
20 V 20 V ±112.5 nA
6 2 V 2 V ±112.5 nA
7
0.5 mA 1 mA
20 V 20 V ±1.125 μA
8 2 V 2 V ±1.125 μA
9
5 mA 10 mA
20 V 20 V ±11.25 μA
10 2 V 2 V ±11.25 μA
11
60 mA 120 mA
20 V 20 V ±0.16 mA
12 2 V 2 V ±0.16 mA

Other manuals for Keysight u2722a

Related product manuals