18 U2722A/U2723A Service Guide
3 Performance Verification and Adjustment
Performance Verification and Adjustment
This section contains the performance verification and
adjustment procedures. If the U2722A/U2723A fails any of
the tests or if any abnormal test results are obtained, return
the unit to the Keysight Service Center for readjustment.
Readjustment can also be carried out using Keysight
Measurement Manager revision 1.7 or above.
Recommended Test Equipment
The recommended test equipment for the performance
verification is listed in the table below. If the exact
equipment is not available, substitute the performance
verification standards of equivalent requirement(s).
Test Consideration
For optimum performance, all procedures should comply
with the following recommendations.
Ensure that the ambient temperature is stable and
between 22 °C and 28 °C. The ideal temperature for
performing performance verification should be 25 °C ± 1 °C.
Make sure that the ambient relative humidity is less than
80%.
Allow a three- hour warm-up period before performance
verification.
Keep the test connection cables as short as possible.
Table 3 - 1 Recommended test equipment
Equipment Requirement(s) Recommended model Purpose Used
Digital multimeter
(DMM)
8½ digits resolution Keysight 3458A Measures DC
voltage and
current
Performance
verification and
adjustment