18 U2722A/U2723A Service Guide
3 Performance Verification and Adjustment 
 
Performance Verification and Adjustment
This section contains the performance verification and 
adjustment procedures. If the U2722A/U2723A fails any of 
the tests or if any abnormal test results are obtained, return 
the unit to the Keysight Service Center for readjustment. 
Readjustment can also be carried out using Keysight 
Measurement Manager revision 1.7 or above.
Recommended Test Equipment
The recommended test equipment for the performance 
verification is listed in the table below. If the exact 
equipment is not available, substitute the performance 
verification standards of equivalent requirement(s).
Test Consideration
For optimum performance, all procedures should comply 
with the following recommendations.
• Ensure that the ambient temperature is stable and 
between 22 °C and 28 °C. The ideal temperature for 
performing performance verification should be 25 °C  ± 1 °C.
• Make sure that the ambient relative humidity is less than 
80%.
• Allow a three- hour warm-up period before performance 
verification.
• Keep the test connection cables as short as possible.
Table 3 - 1 Recommended test equipment
 Equipment  Requirement(s)  Recommended  model  Purpose  Used
Digital multimeter 
(DMM)
• 8½ digits resolution Keysight 3458A Measures DC 
voltage and 
current
Performance 
verification and 
adjustment